Skip to main content
Skip table of contents

A-3-2-1 T7 AIN General Specs [T-Series Datasheet]

All specifications are at 25 °C and Vsupply = 5.0 V unless otherwise noted.

ParameterConditionsMinTypicalMaxUnits
Typical Input Range [1]Gain=1-10.6-10.1Volts
Gain=10-1.06-1.01Volts
Gain=100-0.106-0.101Volts
Gain=1000-0.0106-0.0101Volts
Max AIN Voltage to GND [2]Valid Readings-11.5-11.5Volts
Max AIN Voltage to GND [3]No Damage-20-20Volts
Input Bias Current [4]--20-nA
Input Impedance [4]--1-
Max Source Impedance [4]--1-
Integral Linearity ErrorRange=10, 1, 0.1--±0.01%FS
Range=0.01--±0.1%FS
Absolute AccuracyRange=10, 1, 0.1--±0.01%FS
Range=10--±2000μV
Range=1--±200μV
Range=0.1--±20μV
Range=0.01--±0.1%FS
Range=0.01--±20μV
Temperature Coefficient [5]--15-ppm/°C
Channel Crosstalk [6]< 1kHz--100-dB
1kHz - 50kHz-20-dB/dec
High-Speed ADC -3dB Frequency [7]Gain=1, 10-445-kHz
Gain=100-337-kHz
Gain=1000-63-kHz
High-Res ADC -3dB Frequency [7][8]See Note #7,8---
Noise (Peak-To-Peak)See A-3-2-2--<1μV
Effective Resolution (RMS)See A-3-2-2--22bits
Noise-Free ResolutionSee A-3-2-2--20bits

[1] The range is straightforward for single-ended channels. For differential channels you also need to consider common-mode voltage, so see Appendix A-3-2-3 for more details.

[2] This is the maximum voltage on any AIN pin compared to ground for valid measurements on that channel. For single-ended readings on the channel itself, inputs are limited by the "Typical Input Range" above, and for differential readings consult Appendix A-3-2-3. Further, if a channel has over 13.0 volts compared to ground, readings on other channels could be affected. Because all even channels are on one front-end mux and all odd channels on a second front-end mux, an overvoltage (>13V) on a single channel will generally affect only even or only odd channels.

[3] Maximum voltage, compared to ground, to avoid damage to the device. Protection level is the same whether the device is powered or not. This specification is continuous. For brief transients, such as ESD, the level of protection is much higher.

[4] The key specification here is the maximum source impedance. As long as your source impedance is not over this value, there will be no substantial errors due to impedance problems. For source impedance greater than this value, more settling time might be needed.

[5] Accuracy specs on this page are at room temperature, so Tempco is provided as a typical value reflecting how analog input readings change as temperature changes. For applications attempting better accuracy across varying temperatures there are a couple common strategies. One, on some unused AIN acquire a signal that does not change as the T7 temperature changes, and thus is at a known value so you know how much the T7 readings have changed at any time. Two, evaluate your T7 to determine the relationship of error versus temperature across the temperature range of interest.

[6] Typical crosstalk on a grounded AIN pin, with 20Vpp sine wave on adjacent AIN pin. An adjacent AIN pin refers to multiplexer channel location not channel number, e.g. AIN0-AIN2 or AIN1-AIN3 pairs. An adjacent pin is the worst case. This spec is based on crosstalk seen on a grounded AIN pin, but the same applies if any properly driven signal is connected.

[7] This is the bandwidth of the analog hardware. Any frequencies less than this will go through the analog system to the ADC and be part of the digitized waveform. For DC measurements this is of little concern as ResolutionIndex and averaging can be used to get rid of extra noise. For AC measurements, frequency components below the nyquist point can be removed after digitizing, but frequency components above the nyquist point must be removed before digitizing as they will alias. If unwanted signals with frequencies between the nyquist point and analog cutoff frequency are expected, and they are expected to have sufficient magnitude to be above the acceptable noise level, then an external hardware filter must be used (often called an anti-alias or anti-aliasing filter).

[8] The fixed -3dB frequencies from note 6 apply to the high-speed ADC (ResolutionIndex = 1-8), but the high-resolution ADC on the T7-Pro (ResolutionIndex = 9-12) has filtering at much lower frequencies. The frequency response at ResolutionIndex=12 is shown in Figure 22 of the AD7190 datasheet. For the response at ResIndex 9/10/11 multiply those x-axis values by 47.9/12.0/2.4. Figure 22 only shows up to 150 Hz, but know that all higher frequencies are also filtered out, except for a narrow passband at 307 kHz. The width of this passband is about 200 Hz at ResIndex=12 increasing to about 10000 Hz at ResIndex=9.

See also: T7 Noise and Resolution

JavaScript errors detected

Please note, these errors can depend on your browser setup.

If this problem persists, please contact our support.