2.9.1.4 - Duty Cycle Measurement (Mode 4)
Records the high and low time of a signal on the external pin, which provides the duty cycle, pulse width, and period of the signal. Returns 4 bytes, where the first two bytes (least significant word or LSW) are a 16-bit value representing the number of clock ticks during the high signal, and the second two bytes (most significant word or MSW) are a 16-bit value representing the number of clock ticks during the low signal. The clock frequency is determined by TimerClockBase/TimerClockDivisor.
The appropriate value is updated on every edge, so a read returns the most recent high/low times. Note that a duty cycle of 0% or 100% does not have any edges.
To select a clock frequency, consider the longest expected high or low time, and set the clock frequency such that the 16-bit registers will not overflow.
Note that the minimum measurable high/low time is limited by the edge rate limit discussed in Section 2.9.2.
When using the LabJackUD driver the value returned is the entire 32-bit value. To determine the high and low time this value should be split into a high and low word. One way to do this is to do a modulus divide by 216 to determine the LSW, and a normal divide by 216 (keep the quotient and discard the remainder) to determine the MSW.
Writing a value of zero to the timer performs a reset. After reset, a read of the timer value will return zero until a new edge is detected. If a timer is reset and read in the same function call, the read returns the value just before the reset. The duty cycle reset is special, in that if the signal is low at the time of reset, the high-time/low-time registers are set to 0/65535, but if the signal is high at the time of reset, the high-time/low-time registers are set to 65535/0. Thus if no edges occur before the next read, it is possible to tell if the duty cycle is 0% or 100%.
DAQ Devices
- UD Series (U3, U6, UE9)
- U3
- Start with the UD Driver
- User's Guide
- 1 - Installation on Windows
- 2 - Hardware Description
- 2.1 - USB
- 2.2 - Status LED
- 2.3 - GND and SGND
- 2.4 - VS
- 2.5 - Flexible I/O (FIO/EIO)
- 2.6 - AIN
- 2.7 - DAC
- 2.8 - Digital I/O
- 2.9 - Timers/Counters
- 2.9.1 - Timer Mode Descriptions
- 2.9.1.1 - PWM Output (16-Bit, Mode 0)
- 2.9.1.2 - PWM Output (8-Bit, Mode 1)
- 2.9.1.3 - Period Measurement (32-Bit, Modes 2 & 3)
- 2.9.1.4 - Duty Cycle Measurement (Mode 4)
- 2.9.1.5 - Firmware Counter Input (Mode 5)
- 2.9.1.6 - Firmware Counter Input With Debounce (Mode 6)
- 2.9.1.7 - Frequency Output (Mode 7)
- 2.9.1.8 - Quadrature Input (Mode 8)
- 2.9.1.9 - Timer Stop Input (Mode 9)
- 2.9.1.10 - System Timer Low/High Read (Modes 10 & 11)
- 2.9.1.11 - Period Measurement (16-Bit, Modes 12 & 13)
- 2.9.1.12 - Line-to-Line (Mode 14)
- 2.9.2 - Timer Operation/Performance Notes
- 2.9.1 - Timer Mode Descriptions
- 2.10 - SPC (… and SCL/SDA/SCA)
- 2.11 - DB15
- 2.12 - U3-OEM
- 2.13 - Hardware Revision Notes
- 3 - Operation
- 4 - LabJackUD High-Level Driver
- 5 - Low-level Function Reference
- Appendix A - Specifications
- Appendix B - Enclosure and PCB Drawings
- Frequency List for U3 Timer Mode 7
- UD Driver for Windows with U3 Quick Reference
- U3 Base Class for the Exodriver
- U3 Firmware Revision History
- U6
- UE9
- Software Options
- U3
- U12
- T Series
All Support Guides
Search
Shopping Cart
Testimonials
-
Thank you for the quick response. :) Much appreciated.
—Robert, University of Missouri

Comments
No comments yet. Speak up. We're listening.