LJTIA Thermocouple Troubleshooting
Following is a basic procedure for troubleshooting and testing thermcouples connected to an LJTick-InAmp. We will look at the voltages using the test panel in LJControlPanel. This takes software complications out of the loop. Make sure all other LabJack related software is closed.
Start by removing all signals from the LJTIA, and jumper all 4 inputs (INA+/INA-/INB+/INB-) to GND. Turn off all switches on the LJTIA, except #5, to give you a gain of x1 and offset of 0.4 volts. You should see that both channels measured by the LabJack measure about 0.4 volts.
Now turn on switches 3 and 9 (in addition to 5), to give a gain of x51 and offset of 0.4 volts. The voltages measured by the LabJack will likely change a little, but should both still be near 0.4 volts.
Now remove the jumpers, connect the +/- of a thermocouple to INA+/INA-, and also connect a 10k resistor from INA- to GND. Pull on both leads going into INA- to make sure both are clamped securely.
A thermocouple gives you a voltage related to the difference in temperature between the 2 ends. So if the remote end is at the same temperature as the LJTIA end, the voltage difference created by the thermocouple should be about 0. That means you should get a reading about the same as when you had jumpers to ground.
No try changing the temperature of the remote end of the thermocouple. Here are charts where you can look to see what voltage you expect:
http://www.temperatures.com/tctables.html
For example, say you have a Type K thermocouple, the LJTIA end is at 25 degrees C, and the remote end is at 35 degrees C between your fingers. Look at the following table:
That means the difference is +10 degrees C, and from the following table the voltage you will get is about 397 uV:
http://www.pyromation.com/downloads/data/emfk_c.pdf
The voltage at the 25 deg C end is 1.000 mV and the voltage at the 35 deg C end is 1.407 mV, so the difference is 407 uV. That is the voltage difference that should be created across the INA+/INA- terminals. You can confirm it with a DMM if you have one good enough. The LJTIA multiplies this by x51 and adds about 0.4, so your voltage measured on the test panel should be about 0.4208 volts. That is, about 0.0208 volts higher than the reading with no temperature difference.
You can confirm that 0.421 volts by measuring FIO0 versus GND with a DMM and see if it is the same as the test panel reading. If not, the reason is usually that the signal is varying. The DMM will give an average reading over 0.5 or 1.0 seconds, whereas the U3 gives a reading sampled over perhaps 500 microseconds.
Accessories and Sensors Datasheets and Examples
- CB15 Datasheet
- CB25 Datasheet
- CB37 V1.2 Datasheet
- CB37 V2.1 Datasheet
- EB37 Datasheet
- EI-1022 Datasheet
- EI-1034 Datasheet
- EI-1040 Datasheet
- EI-1050 Datasheet
- EI-1050 Utility for Windows
- IDCA-10 Datasheet
- LJTick-CurrentShunt Datasheet
- LJTick-DAC Datasheet
- LJTick-DAC Testing Utility
- LJTick-Divider Datasheet
- LJTick-InAmp Datasheet
- LJTick-Proto Datasheet
- LJTick-RelayDriver Datasheet
- Mux80 Datasheet
- PS12DC Datasheet
- RB12 Datasheet
- RB16 Datasheet and Schematic
All Support Guides
Search
Shopping Cart
Testimonials
-
I was extremely impressed with your prices, the speed and competence of your service, your detailed documentation, and the otherwise competitive technical specs. Compared to your competitors, your website felt very transparent - it took relativley little time to well acquaint myself with the details of your products. [...] I continue to be pleased with your service and with your product.
—Chris, Virginia Tech

Comments
No comments yet. Speak up. We're listening.