LJStreamUD V1.09

LJStreamUD V1.09 has been posted.  Go to the LJStreamUD support page.

Comments

#1

the file header shows a U9 device despite I use a U6 device.The range shown is not correct

13, CH+=13, CH-=199,
UE9 Range=+/-5, y=n

14, CH+=14, CH-=199, UE9 Range=+/-5, y=o

15, CH+=193, CH-=199, UE9 Range=+/-5,
y=p

Each value is stored twice.

Channel 0 is listed as v0 in colum B and as yo in colums R

Channel 1 is listed as v1 in colum C and as y1 in colums S,

and so on...

 

#2

I see the mistakes in the header and will fix in LJStreamUD V1.10.

The reason you see channel 0 listed twice is that v0 is the raw voltage and y0 is the scaled value.  If your scaling equation is y=a then these columns will be identical.

#3

Hello,

I am a new user of the U6. I need all 14 AI channels (single ended) and 16 DI with am max sampling rate of at least 1000 Hz per channel.

When I start LJstreamud 1.06 it shows an normal number of the internal temperature (295 deg) and nonsens on the channel of the digital inputs (-10.457.. ). When I start streaming, the internal temperatur jumps to 429 deg and the DI shos the correkt value.

What does the number 199 in the colums -CH mean?

 

 

#4

When LJStreamUD is not streaming it is acquiring data in command/response mode.

Channel 193 currently only works in stream mode, but will update soon with command/response support.

The U6 internal temperature sensor does not work in stream mode.  It takes too long to settle.

#5

Thank you very much for your answer. So my U6 works correct. I am very happy how it works on my test bench.